EEG Pre-Experiment Checklist

We strongly recommend completing this checklist before you start collecting EEG data. This will ensure you get the best out of your Explore system.

1. Install software

  1. Install the software appropriate for your needs. An overview of our open-source APIs can be found on the software page of our website or in our GitHub Repo.
  2. Regardless which software you use to record with, you will need explorepy to verify impedances. To download explorepy, please consult our explorepy Installation Guide, or download our Windows installer.

2. Set-up equipment

  1. Clip your electrodes into your desired 10-10 positions on the EEG cap. Use the cables we provide to connect the Explore device to the electrodes. Ensure there is no tension on the cables as this can cause the electrodes to disconnect. Reposition your Explore device to alleviate cable tension if needed.
  2. Thread the electrode cables through the cap’s larger holes to reduce cable movement artifacts (we’ve written a blog post on reducing signal noise when carrying out experiments, have a read!).
  3. Attach the Explore device to the EEG cap using velcro stickers / straps / clips. You can pull straps through the small latch on the back of the device.
  4. Ensure that the EEG cap is pulled over your participant’s head correctly. Their ears should be sticking through the cut outs and the chin strap should close comfortably over or under the chin. In sum: the cap should feel tight but comfortable. Please consult our Cap Size Guide for more.

3. Place the REF electrode

  1. Position: For most EEG applications, we recommend placing the REF on the mastoid (TP9 or TP10). This will minimize artifacts from non-relevant brain activity, EMG, and ECG activity.
  2. Application: Prepare the mastoid location with alcohol and abrasive gel for clean skin. Either tuck the REF electrode under the EEG cap behind the ear (you can use tape here), or use the TP9 or TP10 positions in the EEG cap.
    • Gel electrodes: Apply a generous amount of conductive gel onto the REF electrode.
    • Dry electrodes: Apply some disinfectant onto the electrode. If you have trouble lowering overall impedances, you can apply a little conductive gel or paste to the REF electrode.

4. Check impedances

  1. The lower the impedances, the higher the signal quality. Verify good impedances before recording. Use the impedance function in explorepy. For more on this function, please watch our video guide, or see our explorepy documentation.

  2. If you need to improve your impedances, please have a read of our Improving Impedances page.
  3. Once impedances are as desired, end the impedance function (Ctrl + C in command prompt/Python console).

You are now ready to start recording EEG data. For information on how to visualize and record data using ExplorePy: documentation and videos.

Copyright © 2022 Mentalab